Name |
# |
Type |
Description |
Channel Number, TMRM |
1 |
integer |
Channel number associated with the image, as appears in the Multi-Dimensional Open Dialog. Use Edit/Rename to view or change channel number. |
Channel Number, FLIPR |
2 |
integer |
Channel number associated with the image, as appears in the Multi-Dimensional Open Dialog. Use Edit/Rename to view or change channel number. |
Spectral Unmix Coefficient Matrix |
3 |
string |
To determine spectral crossbleed coefficients use the Tools/Calculate Crossbleed Correction Factor main menu point, or the Math/Blind Spectral Unmix with NMF function. See more on the layout of the Spectral Unmix Coefficient Matrix in the description of Math/Spectral Unmix. |
Image stabilizer compares to first frame |
4 |
boolean |
“Yes” is the fastest way of image stabilization. Turn it off if the image changes a lot during the time lapse, and it is not possible to compare later frames to the first frame any more. |
Image stabilizer period (frames) |
5 |
integer |
1 if comparing to the first frame. Use more frames if not comparing to the first frame and the view field drifts slowly. |
ROIs: Cell diameter (pixels) |
6 |
integer |
Diameter of the cell in pixels in the combined TMRM+FLIPR projection image. Cells ranging around this size will be selected. |
ROIs: Margin around cells (pixels) |
7 |
integer |
Cell boundaries will be extended by this amount. |
ROIs: Debris cutoff (percentile) |
8 |
real |
This percentile of the image histogram sets the intensity value where the maximum of the Look Up Table (LUT) is scaled. Use -1 to override this with fixed value set below at "Max value". |
ROIs: Minimum cell fluorescence (%) |
9 |
real |
Cells dimmer than this in filtered rescaled TMRM+FLIPR projection images will be rejected. Increase this value if debris dimmer than the cells is detected. |
ROIs: Sensitivity of cell detection (% of max fluorescence) |
10 |
real |
For "Bound locally", the boundary of each object is determined at this % of the maximal intensity of the object relative to its neighborhood. For "Bound uniformly" this is a pixel intensity value. |
Membrane potential calibration action |
11 |
string |
Select an action to be performed by the Membrane Potential Calibration Wizard |
Calibration configuration file name (*.ips) |
12 |
string |
Configuration file previously saved from the Membrane Potential Calibration Wizard |
Output Excel Data save file name (*.xlsx) |
13 |
string |
Excel Data window output will be created and saved using this name. |